Surgical instruments
Surgical instruments and unit
Systematic dental surgery
Surgery of the bone tissue - a highly sensitive area: the KaVo surgery system has the right tools for many surgical operations: reliable, powerful, safe and economical.
- Easily dismantled for cleaning, simple hygiene and maintenance.
- Two-part design: Compatibility with all INTRA heads ensures very safe and easy work, saving of time and simple replacement.
- Self-locking safety collar for head change.
- Easy and heat-free running even under continuous load owing to stainless steel alloy: long life despite between-patient sterilization.
- Softly knurled for pleasant grip and easier surface hygiene.
- Anatomical shape.
- Balanced weight.
- Fully sterilizable up to 135 °C and thermally disinfectable.
|
|
KaVo Dental GmbH Bismarckring 39 88400 Biberach/Riß Fon: +49 73 51 / 56-0 Fax: +49 73 51 / 56-14 88 info@kavo.com |
The economic solution for all applications requiring a bone block
bone grafts 3rd molars periapical abcess The clinical advantages are tremendousno boneloss very fast healing stable clinical situation after repositioning of the bone
block \="" />
Implant surgery excellence
Precise programming of 10 programmes with up to 10 steps each
Multi-implant: Saves data of up to 11 implants
Multi-user: Personalised user settings and programmes
Bidirectional, wireless data transfer with PC
INTRAsurg 1000 Air: Rotational and reciprocal bone preparation in conjunction with SONICflex bone
Programmed operation
Every dental implant procedure has a surgical sequence for which the individual steps are planned thoroughly in advance.
So doesn’t it make perfect sense to program the settings for these sequences in detail beforehand so they can be called up whenever they are needed?
For when you need everything to go according to plan. With KaVo INTRAsurg 1000 / 1000 Air you can store programmes for various different surgical procedures, including sinus lift with implant placement, bone splitting with implant placement or implanting in accordance with the manufacturer's instructions for the implant system being used.
Alternative mode
You can also use all functions, such as data communication or reciprocal bone treatment, in manual mode – i.e. without any configured programme.
Technical data
Name
Value
Voltage
100 V bis 240 V
LUX motor
INTRA-LUX SL 550
Light: 25,000 Lux
Speed range
300 – 40.000 U/min-1
Max. torque
300 bis 12.000 U/min-1 - 5,5 Ncm
12.000 - 40.000 U/min-1 - 5,5-2,0 Ncm
Pump delivery
20 - 110 ml/min
10 programmes with up to 10 steps,
customisable parameters
- Straight or contra-angle handpiece gearbox ratio
- Max. speed
- Max. torque
- Pump delivery (0-6)
- Reverse/forwards mode for motor
Implantology brilliance
KaVo INTRAsurg 1000 / 1000 Air is equipped with the proven, powerful motor INTRA-LUX SL 550. The available straight and contra-angle handpieces with illumination meet all implant surgery requirements and their attachment is automatically detected by the motor.
Naturally, SONICflex is also equipped with this important feature, bringing light to even the most remote areas of the mouth. The perfect accessory for precision dental work.
A complete implant procedure can thus be carried out; from preparation of the implant cavity to fixing of the implant.
Special motor
Automatic matching
Max. speed in approx. 1 second
Ax. torque preselectable; acoustic signal on reaching 90%; cut-out at 100%
1000 sterilization cycles without maintenance
Open system (no closure of the motor for sterilization)
Contra-angle handpiece INTRA C3-09 (without light)
Total transmission 27:1
Surgery reducing head C3 (Transmission 3:1)
with Surgery reducing shankl C09 (Transmission 9:1)
External and internal cooling (Kirschner and Meyer)
Max. torque: 55 Ncm
Max. operating speed: 11-1,500 rpm
Wide range of use: from smoothing the alveolar crest to screwing in the implants
Provides sufficient speed and power for milling and drilling
At low speeds, always sufficient force for cutting threads and mounting implants
Convincing price/performance ratio
Very high load capacity
Improved vision of the operating area through 25% smaller diameter
All information at a glance
Only 5 keys
Unit can be operated completely using the foot control
Key parameters
- Max. torque
- Max. speed
- Coolant pump setting
- Transfer ratio of the handpiece and contra-angle handpiece
- Direction of motor rotation
Technical data
Unit
Weight
6.9 kg
Size
31 x 23 x 12,5 cm
Input voltage
100 V; 120 V; 230 V
50/60 Hz
Irrigation function
Motor is switched off, only the pump runs
Display in standby state
Speed range; motor direction;
gear step-down of the instrument;
max. torque; pump power (0...6)
Display during use
Actual speed,
actual torque
Foot control
Motor on/off;
Speed (min. - max.);
Clockwise/counterclockwise operation;
Pump on/off;
Irrigation function;
Switch on program;
Select program step (forwards/back)
INTRA S 550 motor
Technology
Collectorless induction motor
Motor speed
Min. 300 rpm;
Max. 40,000 rpm
Torque
5,5 Ncm: 300 –12,000 rpm
5,0 – 2,0 Ncm: 12,000 – 40,000 rpm
Output power
100 W
Software
Program
1 program with 6 steps
Program functions
Gear step-down of the instrument 1 2 to 1201
Max. speed 80,000 rpm
Max. torque 5.5 to 55 Ncm
Pump power 0 to 6
Motor direction
Clockwise/counterclockwise
Torque
ax. torque preselectable; acoustic signal on reaching 90%; cut-out at 100%
Indications
For approximal enamal reduction (AER)For creating space to correct uneven or misaligned teethFor creating space to insert implants To avoid extractions in young persons who have insufficient space in their mouths for permanent teeth
Prof. Joost-Brinkmann, Charité
The implant cleaning tip
Clinical requirement: Subgingival
cleaning of the smooth neck areas of predominantly titanium
implants. Special tips are therefore required when professionally
cleaning the teeth of patients with implants. Conventional,
tempered steel scaler tips are hard and may damage the soft
titanium on the surface, which encourages a swift build-up
of plaque.
KaVo O-drive set
KaVo O-drive OD30 instrument OS wheels (oscillating segmental wheels)
You will always retain a clear overview of what you are doing, even in areas which are difficult to see, thanks to the optimum illumination provided by the KaVo SONICflex 2003 L.
The SONICflex 2003 L handpiece is connected to the INTRAsurg 1000 Air via the air instrument hose.
Four-edged tip (tip no. 80)
Indications:
Removal of root residue
Extraction prior to implantation
Osteotomy (bone splitting)
Setting on the SONICflex: Level 3
Diamond burr (tip no. 81)
Indication:
Preparing the bony window for an external sinus lift
Setting on the SONICflex: Level 3
Diamond-free burr (tip no. 82)
Indication:
Atraumatic removal of the Schneider´s membrane in a sinus lift
Setting on the SONICflex: Steps 2-3
SONICflex bone tips
The SONICflex Bone Saw
(tip no. 87)
The SONICflex Bone Membrane Disc
(tip no. 85)
The SONICflex Bone Harvester
(tip no. 86)
SONICflex bone (tip no. 87)
Indication
Osteotomy
Ideal to allow precise, extremely narrow incisions, when cutting ligaments
SONICflex bone (tip no. 85)
Indication:
Ideal for reliably and gently repositioning the Schneider membrane during sinus augmentation
SONICflex bone (tip no. 86)
Indication:
Ideal for easily and quickly harvesting large quantities of bone during sinus augmentation
INTRAsurg 300 plus
Light motor and light instruments
Automatic instrument recognition
Storage of the torque
Special motor
Automatic matching
Max. speed in approx. 1 second
Ax. torque preselectable; acoustic signal on reaching 90%; cut-out at 100%
1000 sterilization cycles without maintenance
Open system (no closure of the motor for sterilization)
Contra-angle handpiece INTRA LUX CL3-09 (with light) Total transmission 27:1
Surgery reducing head CL3 Transmission 3:1Cooling:: External and internal cooling (Kirschner and Meyer)
Max. torque: 55 Ncm Surgery reducing shank CL09 Transmission 9:1 Max. operating speed: 11-1,500 rpm
(mit INTRAsurg 300 plus)
Light power: 25,000 Lux
Wide range of use: from smoothing the alveolar crest to screwing in the implants.
Provides sufficient speed and power for milling and drilling.
At low speeds, always sufficient force for cutting threads and mounting implants.
Convincing price/performance ratio.
Very high load capacity.
Improved vision of the operating area through 25% smaller diameter.
Rotational + reciprocal = optimum action
Motor-driven rotational bone procedures are quick and powerful.Air-driven
reciprocal bone preparation is precise, controlled and gentle on tissue
and nerves; a very useful choice to have within implant surgery.In
an ideal world, it would be easy to switch between the two modes. And
with KaVo INTRAsurg 1000 Air, it is; only when used in combination can
the two modes really show their strengths to full advantage.
For a sinus lift, for instance, the bone window can now be pre-prepared
rotationally and completed reciprocally. This significantly reduces the
risk of injuring the Schneider's membrane, which can be subsequently
detached from the bone using the appropriate tip. The user then
switches the motor back to rotational action to prepare the implant
cavity. The combination of these technologies ensures shorter treatment
times and reduces risk and trauma for the patient.
Intelligent motor control
Speed: 300 - 40,000 rpm.
Torque: 300 - 12,000 rpm at 5.5 Ncm.
Max. torque 12,000 - 40,000 rpm at 2.0 - 5.5 Ncm.
Reduced torque when not necessary and therefore less heating of the motor and less stress on the bone.
The economical start for future implantologists
Very simple operation
Intelligent motor control
Program with 6 steps
Display of actual speed and actual torque
New straight and contra-angle handpieces
Use
The advantage of this tip is that you only need one instrument to release the membrane.
Hold the tip against the membrane of the prepared bony window and start to carefully move it inwards.
Continue in this way to dig beneath the membrane and release it from the bony wall of the sinus as far as is required for the sinus lift. Then luxate the membrane with the bone lid in a cranial direction.
Use
Following exposure of the facial wall of the maxillary sinus, use the diamond burr to prepare the bony window (do not exert pressure).
If the osteons are very thick, we recommend using the rotating diamond burr for preliminary treatment of the outer compacta. Then use the diamond burr to prepare the bone near to the membrane.
Use
Remaining root
Use the tip like a periotome: Work the root residue in circular movements, digging deeper with the four-edged tip. This will expand the alveoli and cut through the Sharpey’s fibres.
Non-invasive extraction prior to implantation
When treating teeth with multiple roots, you need to separate the roots carefully using a rotating instrument. Once you have done this, work around each root digging downwards in circular movements.
Osteotomy
Place the tip at a slight angle (as when screwing) on the surface of the bone (do not exert pressure), and move backwards and forwards in the required gap path. If the bone you are treating is very hard, use a thin rotating Lindemann burr to remove the outer compacta around the osteotomy gap.
The high oscillation amplitude facilitates effective and precise incisions. It can be set to 120, 160 or 240 µm easily using the ring switch on the KaVo SONICflex 2003 L handpiece. The elliptical motion of the tips means that you can work in any direction. The SONICflex oscillates at 6 kHz and thereby minimises the overheating of the surrounding tissue.
To simplify surgery, a program can be preprogrammed (rotating and oscillating) for the KaVo INTRAsurg 1000 Air. You can easily switch between the programs using the foot control.
Technical data
KaVo O-drive OD 30Gearbox ratio: 7.4:1
Movement: Reciprocal, 30° oscillation
Bur shaft: Multiple-spline shaft
Perfect for reciprocal mechanical action
Minimal space required
Perfect control
Fits securely in the contra-angle handpieceSegmental wheelsKomet, Brasseler: Patent DE 197 54 879OS1 M.000.140 (0.30 mm thick)
OS1 F.000.140 (0.15 mm thick)
OS2 M.000.140 (0.45 mm thick)
OS1 F.000.140 (0.30 mm thick)
Description:
OS = Oscillating stripping
F = Medium diamond coating D54
M = Fine diamond coating D 25
000 = Special shaft
140 = Wheel radius (14 mm here)
The solution - SONICflex Implant
For the effective removal of subgingival plaque and concretion No abrasion, no scratching Quick and easy to useA patient product: effective, no risk of infection, excellent price/performance ratioProphylactic implant cleaning stages Runs on > sonicflex="" 2003="" (l)="">SONICflex 2003 (L) , level 1
1. Tooth width measurement
Features
Faster and more precise than manual enamel stripping No damage to soft tissue as a result of reciprocal action Radial movement over a large tooth areaEnamel is reduced in one go: No stages, no tilting - quick procedure
Speeds and torques in implantology
Clinical procedure
Speed (rpm)
Max. torque required (Ncm)
Smoothing the alveolar crest
1,000 - 1,500
5 - 10
Drilling the cavity
700 - 1,500
till 15
Thread cutting
- Standard implants
- Large diameter
10 - 50
10 - 50
till 35
till 50
Screwing in implant
- Standard implants
- Large diameter
10 - 50
10 - 50
till 35
till 50
Prosthetics
10 - 50
till 35
INTRAsurg 300 and C3-09
11 - 1,500
till 55
Program with 6 steps
Step
Procedure
Instrument
Max.
speed
Torque
Pump
Motor direction
1.
Milling and drilling the implant bed
27:1
1.500
55 Ncm
4
2.
Cutting the thread
27:1
50
55 Ncm
3
3.
Cutting instrument back
27:1
50
55 Ncm
3
4.
Screwing in implant
27:1
50
55 Ncm
3
5.
Removing the insertion post
27:1
50
55 Ncm
0
6.
Mounting the closure cap
27:1
50
55 Ncm
0
The procedure for an implant operation is shown.
For simplification, the essential parameters are specified but you can change them all individually.
Intelligent motor control
Speed: 300 - 40,000 rpm
Torque: 300 - 12,000 rpm at 5.5 Ncm
Max. torque 12,000 - 40,000 rpm at 2.0 - 5.5 Ncm
Reduced torque when not necessary and therefore less heating of the motor and less stress on the bone
Intelligent motor control
Speed: 300 - 40,000 rpm.
Torque: 300 - 12,000 rpm at 5.5 Ncm.
Max. torque 12,000 - 40,000 rpm at 2.0 - 5.5 Ncm.
Reduced torque when not necessary and therefore less heating of the motor and less stress on the bone.
Intelligent motor control
FeaturesSpeed: 300 - 40,000 rpmTorque:
300 - 12,000rpm = 5.5 Ncmmax. torque, if required12,000 - 40,000 rpm = 2.0 to 5.5 NcmReduced torque, if not required, thus motor heats up less and bone is subject to less strain
The economical start for future implantologists
Very simple operation
Intelligent motor control
Program with 6 steps
Display of actual speed and actual torque
New straight and contra-angle handpieces
Additional INTRAsurg 1000 Air equipment
Air instrument connection (for reciprocal mechanical action)2nd peristaltic pump Pneumatic tubing (not sterilisable) External irrigant tubing (sterilisable) MULTIflex coupling
The configuration is always right
Before surgery, you preselect the treatment parameters you need for the procedure in question, and INTRAsurg 1000 /1000 Air adheres to these during the procedure. The customisable parameters are as follows:
Rotational or reciprocal mechanical action
(INTRAsurg 1000 Air)
Torque
Speed
Gearbox ratio
Cooling
Rotational direction of motor
Surgery reducing head 67 LHB 2:1For diamond grinding discs in combination with
push fit disc guard. Disc guard, short/ longFor discs ISO 204 080/205 080.
Foot control
The foot control for KaVo INTRAsurg 1000 / 1000 Air, surgically approved (IPX8)
Surgery shank 3620 N 1:1Easily dismantled. Optimal hygiene.
Internal coolant tubing
No inconvenient manipulation
Very flexible
Easily changeable
Sterilizable; sterile disposable tube fittings available
How to use the segmental wheel
Operating range: 2,700 to 5,500 oscillations per minute Place the segmental wheel from the occlusal or vestiblar and direct through the point of contact in a slow and steady movement. Only use if there is sufficient spray cooling (at least 50 ml per min.) This procedure must be followed by an approximal polish
2. Teeth are stripped from occlusal to vestibular in a single operation (with 2,700 - 5,500 U/min)
Features
SONICflex clean tip no. 48 Plastic base (PEEK)Single useNo cross-infection
With KaVo’s SONICflex bone set, you can easily access new and minimally invasive forms of treatment.
The KaVo SONICflex bone set has 5 tips for all manner of applications. It can be used for oscillating osteotomies (e.g. distraction) and bone splitting, as well as for the minimally invasive removal of teeth and root residue. Moreover, it increases safety in the context of external sinus lifts.
Case documentation
The series of images shows an example of how the four-edged tip can be used for the minimally invasive extraction of a hemisectioned tooth with apical foci.
Pictures: Dr. F. Heinemann, Morsbach
Case documentation
The pre-operative x-ray reveals the extremely restricted amount of vertical bone in the left posterior maxilla.
Since a subsequent implantation was intended, an external sinus lift was performed to increase the amount of bone.
In this case, a rotating instrument was used for preliminary treatment. The remaining bone around the osteotomy line was then removed and smoothed with the diamond burr.
Pictures: Dr. Dr. S.-H. Gnoth, Cologne
Case documentation
Once the bony sinus window has been prepared, the Schneider’s membrane is carefully removed. The membrane was completely undermined in the caudal region and luxated in a cranial direction.
The subantral space gained by means of this process is filled with bone substitute and sealed with a resorbable membrane.
The post-operative x-ray reveals the significant lift of the sinus floor by means of the augmentation (the pre-operative X-ray appears above in the section describing the diamond burr tip).
Pictures: Dr. Dr. S.-H. Gnoth, Cologne
Axial and sagittal saws
Indications:
Removal of bone blocks
Osteotomy
Setting on the SONICflex: Level 3
Internal coolant tubing
No inconvenient manipulation
Very flexible
Easily changeable
Sterilizable; sterile disposable tube fittings available
KaVo SONICflex bone; for reciprocal mechanical action
Tried-and-tested working methodLarge, variable swing arc
120, 160 and 240 µm Particularly effective and extremely precise Low frequency (6,000 Hz)
Minimises heat generation No risk of necrosis Elliptical tip movement
Offers complete freedom of movement - both horizontal and vertical Ultra quiet
Surgical contra-angle handpiece INTRA LUX CL3-04
The fast implantology contra-angle handpiece CL 3-04 with a reduction ratio of 12:1. For implant systems with fast pilot drills up to 3,000 rpm.
Features:
Cooling: External and internal cooling (Kirschner and Meyer) Max. operating speed: 25 - 3,300 rpm Max. torque: 45 Ncm Light output: 25,000 Lux
Technical data
Unit
Weight
6.9 kg
Size
31 x 23 x 12,5 cm
Input voltage
100 V; 120 V; 230 V
50/60 Hz
Irrigation function
Motor is switched off, only the pump runs
Display in standby state
Speed range; motor direction;
gear step-down of the instrument;
max. torque; pump power (0...6)
Display during use
Actual speed,
actual torque
Foot control
Motor on/off;
Speed (min. - max.);
Clockwise/counterclockwise operation;
Pump on/off;
Irrigation function;
Switch on program;
Select program step (forwards/back)
INTRA-LUX SL 550 motor
Technology
Collectorless induction motor
Motor speed
Min. 300 rpm;
Max. 40,000 rpm
Torque
5,5 Ncm: 300 –12,000 rpm
5,0 – 2,0 Ncm: 12,000 – 40,000 rpm
Output power
100 W
Light
25,000 LUX
Software
Program
1 program with 6 steps
Program functions
Gear step-down of the instrument 1 2 to 1201
Max. speed 80,000 rpm
Max. torque 5.5 to 55 Ncm
Pump power 0 to 6
Motor direction
Clockwise/counterclockwise
Torque
ax. torque preselectable; acoustic signal on reaching 90%; cut-out at 100%
Contra-angle handpiece INTRA LUX CL3-09 (with light) Total transmission 27:1
Surgery reducing head CL3 Transmission 3:1Cooling:: External and internal cooling (Kirschner and Meyer)
Max. torque: 55 Ncm Surgery reducing shank CL09 Transmission 9:1 Max. operating speed: 11-1,500 rpm
(mit INTRAsurg 300 plus)
Light power: 25,000 Lux
Wide range of use: from smoothing the alveolar crest to screwing in the implants.
Provides sufficient speed and power for milling and drilling.
At low speeds, always sufficient force for cutting threads and mounting implants.
Convincing price/performance ratio.
Very high load capacity.
Improved vision of the operating area through 25% smaller diameter.
Surgical contra-angle handpiece INTRA LUX CL3-04
The fast implantology contra-angle handpiece CL 3-04 with a reduction ratio of 12:1. For implant systems with fast pilot drills up to 3,000 rpm.
Features:
Cooling: External and internal cooling (Kirschner and Meyer) Max. operating speed: 25 - 3,300 rpm Max. torque: 45 Ncm Light output: 25,000 Lux
Speeds and torques in implantology
Clinical procedure
Speed (rpm)
Max. torque required (Ncm)
Smoothing the alveolar crest
1,000 - 1,500
5 - 10
Drilling the cavity
700 - 1,500
till 15
Thread cutting
- Standard implants
- Large diameter
10 - 50
10 - 50
till 35
till 50
Screwing in implant
- Standard implants
- Large diameter
10 - 50
10 - 50
till 35
till 50
Prosthetics
10 - 50
till 35
INTRAsurg 300 plus
11 - 1,500
till 55
Optimum sterilization
Motor, CL3-09 and tubes completely sterilizable
Very easy and reliable
The comprehensive solution for perfectly hygienic work
All information at a glance
Only 5 keys
Unit can be operated completely using the foot control
Key parameters
- Max. torque
- Max. speed
- Coolant pump setting
- Transfer ratio of the handpiece and contra-angle handpiece
- Direction of motor rotation
Incl. instrument recognition
Contra-angle handpiece INTRA C3-09 (without light)
Total transmission 27:1
Surgery reducing head C3 (Transmission 3:1)
with Surgery reducing shankl C09 (Transmission 9:1)
External and internal cooling (Kirschner and Meyer)
Max. torque: 55 Ncm
Max. operating speed: 11-1,500 rpm
Wide range of use: from smoothing the alveolar crest to screwing in the implants
Provides sufficient speed and power for milling and drilling
At low speeds, always sufficient force for cutting threads and mounting implants
Convincing price/performance ratio
Very high load capacity
Improved vision of the operating area through 25% smaller diameter
Use
Move the saws slowly backwards and forwards in the required gap path, taking care not to exert pressure.
In cases of very dense and hard outer compacta bone tissue, preliminary treatment of the area around the osteotomy gap with a thin Lindemann burr may be necessary.
Proper use of the SONICflex bone tips
Video: Faster and gentler sinus lift with the SONICflex bone
3. Essential: Spray to cool
Optimum sterilization
Motor, C3-09 and tubes completely sterilizable
Very easy and reliable
The comprehensive solution for perfectly hygienic work
Surgery handpiece INTRA LUX CL 10 (with light)
Total step-down ratio: 1:1 Cooling: External cooling (Kirschner
and Meyer) Max. torque: 10 Ncm Max. operating speed: 300-40,000 rpm
(with INTRAsurg 300 plus)Light power: 25000 lux
Surgery insert and chuck.
Surgery disc guard For discs ISO 205 080.
The KaVo power house. Pure power at low speeds Total transmission 120:1
Surgery reducing head
67 RIC Transmission 4:1 (0.540.8340)With internal cooling system (according to Kirschner
and Meyer)and/or external coolant connection. Max. load: 50 Ncm. Surgery reducing shank 3630 N Transmission 30:1 (0.534.5670)Operating speed: 3 - 340 rpm (lowest speed is dependent
on motor, in this case KaVo INTRAsurg 500). Easily dismantled. Improved head chuck. Stronger reducing gear. .
Intuitive operation
Actual speed/torque display
All information at a glance
Only 5 keys
Instrument can be operated completely by foot control
4. The entire tooth is stripped in a single movement
Used in conjunction with the INTRAsurg 1000 Air, SONICflex bone provides you with the ideal combination of rotating and oscillating surgical instruments in a single unit. The power and speed of the proven rotating operating principle is impressive.
In sensitive and critical areas, you can protect a maximum amount of soft tissue and nerves by switching to the minimally-invasive oscillating KaVo SONICflex bone.
Case documentation
This patient was due to undergo an implantation in the area of tooth 11. Due to the buccal bone atrophy, augmentation with a retromolar bone block from the right mandible was undertaken prior to implantation.
Pictures: Dr. E. Guez, Paris
Program with 6 steps
Step
Procedure
Instrument
Max.
speed
Torque
Pump
Motor direction
1.
Milling and drilling the implant bed
27:1
1.500
55 Ncm
4
2.
Cutting the thread
27:1
50
55 Ncm
3
3.
Cutting instrument back
27:1
50
55 Ncm
3
4.
Screwing in implant
27:1
50
55 Ncm
3
5.
Removing the insertion post
27:1
50
55 Ncm
0
6.
Mounting the closure cap
27:1
50
55 Ncm
0
The procedure for an implant operation is shown.
For simplification, the essential parameters are specified but you can change them all individually.
Intuitive operation
Actual speed/torque display
All information at a glance
Only 5 keys
Instrument can be operated completely by foot control with IPX 8
Contra-angle handpiece INTRA LUX CL3-09 (with light) Total transmission 27:1
Surgery reducing head CL3 Transmission 3:1Cooling:: External and internal cooling (Kirschner and Meyer)
Max. torque: 55 Ncm Surgery reducing shank CL09 Transmission 9:1 Max. operating speed: 11-1,500 rpm
(mit INTRAsurg 300 plus)
Light power: 25,000 Lux
Wide range of use: from smoothing the alveolar crest to screwing in the implants.
Provides sufficient speed and power for milling and drilling.
At low speeds, always sufficient force for cutting threads and mounting implants.
Convincing price/performance ratio.
Very high load capacity.
Improved vision of the operating area through 25% smaller diameter.
Surgery handpiece INTRA LUX CL 10 (with light)
Total step-down ratio: 1:1 Cooling: External cooling (Kirschner
and Meyer) Max. torque: 10 Ncm Max. operating speed: 300-40,000 rpm
(with INTRAsurg 300 plus)Light power: 25000 lux
Surgical motor INTRA-LUX SL 550
Techn. Data (Mat.-No. 1.001.3421)
Motorsteuerung
INTRAsurg 300 plus, INTRAsurg 1000, INTRAsurg 1000 Air, treatment unit with electronic
Speed range
min. 300 rpm / max. 40,000 rpm
Torque control
max. 5.5 Ncm
Light
25,000 Lux
Surgery handpiece 3610 N1 1:1 (0.524.5600)
For handpiece bur with standard length 44.5 mm. May also be used for contra-angle burs if the supplied
spacer is employed.
Surgical contra-angle handpiece INTRA LUX CL3-04
The fast implantology contra-angle handpiece CL 3-04 with a reduction ratio of 12:1. For implant systems with fast pilot drills up to 3,000 rpm.
Features:
Cooling: External and internal cooling (Kirschner and Meyer) Max. operating speed: 25 - 3,300 rpm Max. torque: 45 Ncm Light output: 25,000 Lux
Advantages of using the KaVo SONICflex bone and KaVO INTRAsurg 1000 Air
A simple and cost-effective way to expand your range of treatments The ideal combination of rotating and oscillating surgery in a single unit Maximum tissue protection and minimum post-operative trauma Precise and effective incisions Can be used in many areas of bone treatment (osteotomies, augmentation, extractions and sinus lift) Optimum illumination of areas difficult to visualise. Intelligent supply of coolant to the working tip without compressed air With the KaVo INTRAsurg 1000 Air, you can now also use KaVo’s SONICflex paro, rootplaner and retro tips with sterile coolant
5. Procedure complete - the enamel is reduced on both sides
SONICflex 2003
FeaturesLarge, variable swing arc
120, 160 and 240 µm
Particularly effective and extremely preciseLow frequency (6,000 Hz)
Minimises heat generation
No risk of necrosisElliptical tip movement
Offers complete freedom of movement - both horizontal and verticalIntensity
25,000 LUXUltra quiet
Surgery handpiece INTRA LUX CL 10 (with light)
Total step-down ratio: 1:1
Cooling: External and internal cooling (Kirschner and Meyer)
Max. torque: 10 Ncm
Max. operating speed: 300-40,000 rpm
(with INTRAsurg 300 plus)
Light power: 25,000 lux
Automatic instrument recognition
Optimum matching
KaVo INTRAsurg 300 plus recognizes the three light instruments CL3-09, CL3-04 and CL10 immediately
Whether you want to use the straight or the contra-angle handpiece - the right step-down ratio will be set immediately
Surgery handpiece 3610 N2 1:1 (0.524.5610)
For handpiece burs with special length 62.5 mm. May also be used for standard handpiece burs (44.5
mm) if the supplied spacer is employed.
Surgery handpiece 3610 N3 1:1 (0.524.5620)
Only for handpiece burs with special length 70.0
mm. With reinforced drive insert.
SONICflex implant
The implant cleaning tip. Mehr ...
Storage of the torque
At the program level, you are mounting up to six implants in the program steps. With the maximum torque value automatically determined in each step, you can safely assess the primarily stability of the mounted implants. For each implant, you can decide when load can be applied.
Manual level (display 1):
All important settings at a glance - preparation made easy.
Program level (display 2):
Controlled screwing in of the implants.
Program level (display 3):
Display of the maximum torques for assessing the primary stability.
Intelligent motor control
Speed: 300 - 40,000 rpm
Torque: 300 - 12,000 rpm at 5.5 Ncm
Max. torque 12,000 - 40,000 rpm at 2.0 - 5.5 Ncm
Reduced torque when not necessary and therefore less heating of the motor and less stress on the bone
Surgery reducing handpiece 3614 N 4:1 (0.524.5430)
For handpiece bur.
Speed 100 - 10,000 rpm .
Surgical motor INTRA S 550
Techn. Data (Mat.-No. 1.000.8072)
Motor control
INTRAsurg 300, INTRAsurg 300 plus, INTRAsurg 1000, treatment unit with elektronic
Speed range
min. 300 rpm / max. 40,000 rpm
Torque control
max. 5.5 Ncm